Fourth generation of CT scanner uses
Rotate-Fixed Ring geometry where a ring of fixed detectors completely surrounds
the patient. Tungsten alloy X-ray tube rotates inside the detector ring through a full 360
degrees with a wide fan beam producing a single image. Due to the elimination
of translate-rotate motion the scan time is reduced comparable with third generation
scanner, initially, to 10 seconds per slice but the radiographic geometry is
poor because the X-ray tube must be closer to the patient than the detectors,
i.e. the geometric magnification is large also scatter artifact is more than
third generation since they cannot use anti-scatter grid. -The disadvantages of
poor geometry noted above have been alleviated very neatly by the so called
nutating geometry. Tungsten alloy X-ray tube is external to the detector ring but slightly
out of the detector plane, this change resulted in increasing both the
acquisition speed, and image resolution. The method of scanning was still slow,
because the X-ray tube and control components interfaced by cable, limiting the
scan frame rotation. Further, they were more sensitive to artifacts because the
non-fixed relationship to the x-ray source made it impossible to reject
scattered radiation.
没有评论:
发表评论